EasyManua.ls Logo

Rohde & Schwarz RTE - Page 1169

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1169User Manual 1326.1032.02 ─ 20
Usage: Asynchronous command
TRIGger<m>:ANEDge:FILTer <Filter>
Sets a filter for the external trigger signal to reject high or low frequencies.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Filter> OFF | LFReject | RFReject
OFF
The trigger signal is not filtered.
LFReject
Frequencies higher than the "Cut-off" frequency are rejected,
lower frequencies pass the filter.
You can adjust the "Cut-off" frequency using the TRIGger<m>:
ANEDge:CUToff:LOWPass command, the default is 50 kHz.
RFReject
Frequencies below the "Cut-off" frequency are rejected, higher
frequencies pass the filter.
You can adjust the "Cut-off" frequency using the TRIGger<m>:
ANEDge:CUToff:HIGHpass command, the default is 50 kHz.
*RST:
OFF
Usage: Asynchronous command
TRIGger<m>:ANEDge:CUToff:HIGHpass <AnalogCutOffHP>
Frequencies below the "Cut-off" frequency are rejected, higher frequencies pass the fil-
ter.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<AnalogCutOffHP> KHZ5 | KHZ50 | MHZ50
Cut-off frequency
KHZ5
5 kHz
KHZ50
50 kHz
MHZ50
50 MHz
*RST: KHZ50
Usage: Asynchronous command
Trigger

Table of Contents

Other manuals for Rohde & Schwarz RTE

Related product manuals