Remote control commands
R&S
®
RTE
1285User Manual 1326.1032.02 ─ 20
UPPK
The y-value at the maximum sample value in the upper half of
the histogram
LWPK
The y-value at the maximum sample value in the lower half of
the histogram
MAXimum
The highest y-value with a probability > 0
MINimum
The lowest y-value with a probability > 0
MEDian
The y-value for which half the samples lie above, the other half
below in the histogram
MEAN
The weighted arithmetic average of the histogram
*RST: 
PEAK
Usage:  Asynchronous command
MEASurement<m>:HISTogram:PROBability:LIMit <HistgPrbDomLim>
Defines a range around the probability marker.
Suffix: 
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<HistgPrbDomLim> Range:  0  to  100
Increment:  10
*RST:  10
Default unit: %
Usage:  Asynchronous command
MEASurement<m>:HISTogram:LCHeck<n>:VALid <ValidRange>
Enables or disables limit checking for histogram measurements in the specified mea-
surement channel.
Suffix: 
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
<n> 1..18
Number of the histogram measurement type, see Table 18-8.
Parameters:
<ValidRange> ILIMit | ULIMit | LLIMit | OLIMit
Automatic measurements