EasyManua.ls Logo

Rohde & Schwarz RTE - Page 1292

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1292User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<RstWfmCnt> Range: 1 to 65535
Increment: 10
*RST: 1000
Usage: Asynchronous command
MEASurement<m>:STATistics:RMEascount
<RstMeasCnt>
Defines the number of measurement results from which one point of the long term
measurement is created.
This setting is only available if MEASurement<m>:STATistics:MODE is set to MEAS.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<RstMeasCnt> Range: 1 to 65535
Increment: 10
*RST: 1000
Usage: Asynchronous command
MEASurement<m>:STATistics:RTIMe <ResetTime>
Defines the time or period after which the statistics are reset.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<ResetTime> Range: 0.1 to 2.14748E+9
Increment: 1E-3
*RST: 0.2
Default unit: s
Usage: Asynchronous command
MEASurement<m>:STATistics:RESet
Resets the histogram, the long term measurement and the statistics.
Automatic measurements

Table of Contents

Other manuals for Rohde & Schwarz RTE

Related product manuals