Remote control commands
R&S
®
RTE
1337User Manual 1326.1032.02 ─ 20
Setting parameters:
<MaskTestName> String with the name of the mask test
<MaskSegmIdx> Number of the segment. Counting starts from 0.
Usage: Setting only
Asynchronous command
MTESt:SEGMent:REMove <MaskTestName>, <MaskSegmIdx>
Removes the specified segment from the mask definition.
Setting parameters:
<MaskTestName> String with the name of the mask test
<MaskSegmIdx> Number of the segment. Counting starts from 0.
Usage: Setting only
Asynchronous command
MTESt:SEGMent:CLEar <MaskTestName>
Deletes all mask segments of the indicated mask test.
Setting parameters:
<MaskTestName> String with the name of the mask test
Usage: Setting only
Asynchronous command
MTESt:SEGMent:REGion <MaskTestName>, <MaskSegmIdx>,<Region>
MTESt:SEGMent:REGion? <MaskTestName>, <MaskSegmIdx>
Defines the region of the segment that builds the mask.
Parameters:
<Region> UPPer | LOWer | INNer
UPPer
the segment points are connected to a line, the display area
above this line is the mask segment
LOWer
the segment points are connected to a line, the display area
below this line is the mask segment
INNer
the segment points form a closed geometrical shape, which is
the mask segment
*RST: INNer
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
<MaskSegmIdx> Number of the segment. Counting starts from 0.
Mask testing