Remote control commands
R&S
®
RTE
1340User Manual 1326.1032.02 ─ 20
Setting parameters:
<MaskTestName> String with the name of the mask test
<MaskSegmIdx> Number of the segment. Counting starts from 0.
Usage: Setting only
Asynchronous command
MTESt:SEGMent:RESCale:XFACtor <MaskTestName>,
<MaskSegmIdx>,<ExpansFactX>
MTESt:SEGMent:RESCale:XFACtor? <MaskTestName>, <MaskSegmIdx>
MTESt:SEGMent:RESCale:YFACtor <MaskTestName>,
<MaskSegmIdx>,<ExpansFactY>
MTESt:SEGMent:RESCale:YFACtor? <MaskTestName>, <MaskSegmIdx>
Stretches or compresses the selected mask segment in horizontal (XFACtor) or vertical
direction (YFACtor). The x- or y-values of all points of the selected mask segment are
multiplied with this factor. Factors >1 stretch the mask segment, while factors between
0 and 1 compress it. Negative values are possible and change the algebraic sign.
Only takes effect after the MTESt:SEGMent:RESCale:RECalculate command.
Parameters:
<ExpansFactY> Range: -100 to 100
Increment: 1
*RST: 1
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
<MaskSegmIdx> Number of the segment. Counting starts from 0.
Usage: Asynchronous command
MTESt:SEGMent:RESCale:XOFFset <MaskTestName>, <MaskSegmIdx>,<OffsetX>
MTESt:SEGMent:RESCale:XOFFset? <MaskTestName>, <MaskSegmIdx>
Moves the mask segment horizontally. The specified offset is added to the x-values of
all points of the selected mask segment.
Only takes effect after the MTESt:SEGMent:RESCale:RECalculate command.
Parameters:
<OffsetX> Range: -50 to 50
Increment: 1E-9
*RST: 0
Default unit: s
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
<MaskSegmIdx> Number of the segment. Counting starts from 0.
Usage: Asynchronous command
Mask testing