Remote control commands
R&S
®
RTE
1396User Manual 1326.1032.02 ─ 20
MMEMory:SAV <FileDestination>
Stores the current instrument settings to the specified file.
This command has the same effect as the combination of *SAV and
MMEMory:STORe:STATe.
Parameters:
<FileDestination> String parameter specifying path and filename of the target file.
Wildcards are not allowed.
Example:
MMEM:SAV 'C:\mysavefile.dfl'
Saves the current instrument settings to the file
mysavefile.dfl located in the directory C:\.
Usage: Event
MMEMory:RCL <FileSource>
Restores the instrument settings from the specified file.
This command has the same effect as the combination of MMEMory:LOAD:STATe and
*RCL.
Parameters:
'<FileSource>' String parameter specifying the path and filename of the source
file. Wildcards are not allowed.
Example:
MMEM:RCL 'C:\mysavefile.dfl'
Loads and activates the instrument settings from the file
mysavefile.dfl located in the directory C:\.
Usage: Event
MMEMory:STORe:STATe <MemoryNumber>, <FileName>
Stores the instrument settings from the specified internal memory to the specified file.
To store the current instrument settings to the internal memory, use *SAV first.
Setting parameters:
<MemoryNumber> Number of the internal memory
Range: 1 to 99
<FileName> String parameter specifying the complete path and filename of
the source file.
Example:
*SAV 4
MMEM:STORe:STATe 4,'C:
\Settings\Settings_1051.dfl'
Saves current instrument settings to the internal memory num-
ber 4. Then stores the settings from the internal memory number
4 to the file C:\Settings\Settings_1051.dfl.
Usage: Setting only
Data management