Remote control commands
R&S
®
RTE
1080User Manual 1326.1032.02 ─ 20
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RCL <Number>
Recall
Loads the instrument settings from an intermediate memory identified by the specified
number. The instrument settings can be stored to this memory using the command
*SAV with the associated number.
It also activates the instrument settings which are stored in a file and loaded using
MMEMory:LOAD:STATe.
*RST
Reset
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
Usage:
Setting only
*SAV <Number>
Save
Stores the current instrument settings under the specified number in an intermediate
memory. The settings can be recalled using the command *RCL with the associated
number.
To transfer the stored instrument settings to a file, use MMEMory:STORe:STATe.
*SRE <Contents>
Service request enable
Sets the service request enable register to the indicated value. This command deter-
mines under which conditions a service request is triggered.
Parameters:
<Contents> Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range: 0 to 255
*STB?
Status byte query
Common commands