Remote control commands
R&S
®
RTE
1437User Manual 1326.1032.02 ─ 20
BUS<m>:I2C:FRAMe<n>:ADDRess?
Returns the device address value of the indicated frame, that is, the address value that
is shown in the decoded cells and in the decode results table.
If the frame has a 7-bit address, the command considers the status of BUS<m>:I2C:
RWBit. If BUS<m>:I2C:RWBit INADdress is set, the returned address includes the
R/W bit (8 bit). Otherwise, the pure address without the R/W bit is returned (7 bit, same
result as returned with BUS<m>:I2C:FRAMe<n>:ADEVice?).
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<AddressValue> To set the value format, use FORMat:BPATtern.
The values below – range, increment and default – are decimal
values.
Range: 0 to 1023
*RST: 0
Usage: Query only
BUS<m>:I2C:FRAMe<n>:ADEVice?
Returns the pure device address of the indicated frame without the R/W bit.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<DeviceAddress> To set the value format, use FORMat:BPATtern.
The values below – range, increment and default – are decimal
values.
Range: 0 to 1023
*RST: 0
Usage: Query only
BUS<m>:I2C:FRAMe<n>:AMODe?
Returns the address length.
Suffix:
<m>
.
1..4
Selects the serial bus.
Protocols