Remote control commands
R&S
®
RTE
1449User Manual 1326.1032.02 ─ 20
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with SEARch:TRIGger:I2C:DPOSition.
INRange = RANGe
In range: Set the minimum and maximum value of the range with
SEARch:TRIGger:I2C:DPOSition and SEARch:TRIGger:
I2C:DPTO.
*RST: ANY
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:I2C:DPOSition <SearchName>,<DataPosition>
SEARch:TRIGger:I2C:DPOSition? <SearchName>
Defines the first byte of interest. All bytes before that byte are ignored.
Parameters:
<DataPosition> The index 0 is associated with the first data byte.
Range: 0 to 4095
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:I2C:DPTO <SearchName>,<DataPositionTo>
SEARch:TRIGger:I2C:DPTO? <SearchName>
Defines the last byte of interest, if SEARch:TRIGger:I2C:DPOPerator defines a
range.
Parameters:
<DataPositionTo> Range: 0 to 4095
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:I2C:DCONdition <SearchName>,<DataOperator>
SEARch:TRIGger:I2C:DCONdition? <SearchName>
Sets the operator to set a specific data value or a data range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
*RST: EQUal
Protocols