Remote control commands
R&S
®
RTE
1464User Manual 1326.1032.02 ─ 20
*RST: SSACtive
Usage: Asynchronous command
TRIGger<m>:SPI:PALignment <DataAlignment>
Defines how the specified data pattern is searched.
Parameters:
<DataAlignment> WORD | BIT
WORD
The pattern is matched only at word boundaries.
BIT
Bit-by bit: the pattern can be at any position in the data word.
*RST: WORD
Usage: Asynchronous command
TRIGger<m>:SPI:DPOPerator
<DataPosOperator>
Sets the operator for the data position. You can defined an exact position, or a position
range.
Parameters:
<DataPosOperator> ANY | OFF | EQUal | GETHan | INRange | RANGe
ANY = OFF
The position of the required pattern is not relevant.
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with TRIGger<m>:SPI:DPOSition.
INRange = RANGe
Set the minimum and maximum value of the range with
TRIGger<m>:SPI:DPOSition and TRIGger<m>:SPI:DPTO.
*RST: ANY
TRIGger<m>:SPI:DPOSition <DataPosition>
Sets the number of bits or words to be ignored before the first bit or word od interest.
The effect is defined by TRIGger<m>:SPI:PALignment.
Parameters:
<DataPosition> The index 0 is associated with the first data byte.
Range: 0 to 4095 for triggering on one line (MISO or
MOSI), 2047 for triggering on both lines.
Increment: 1
*RST: 0
Protocols