Remote control commands
R&S
®
RTE
1547User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:LIN:WUFRame <SearchName>,<WakeUpFrame>
SEARch:TRIGger:LIN:WUFRame? <SearchName>
Enables the search for wakeup frames.
Parameters:
<WakeUpFrame> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:LIN:ICONdition
<SearchName>,<IdOperator>
SEARch:TRIGger:LIN:ICONdition? <SearchName>
Sets the operator to define a specific identifier or an identifier range.
Parameters:
<IdOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These operators require one identifier
pattern to be set with SEARch:TRIGger:LIN:IMIN
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with SEARch:TRIGger:LIN:IMIN and SEARch:
TRIGger:LIN:IMAX.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:LIN:IMIN <SearchName>,<IdPattern>
SEARch:TRIGger:LIN:IMIN? <SearchName>
Specifies a slave identifier pattern, or sets the the start value of an identifier range.
Parameters:
<IdPattern> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:LIN:IMAX <SearchName>,<IdPatternTo>
SEARch:TRIGger:LIN:IMAX? <SearchName>
Sets the the end value of an identifier range if SEARch:TRIGger:LIN:ICONdition
is set to INRange or OORange.
Protocols