Remote control commands
R&S
®
RTE
1608User Manual 1326.1032.02 ─ 20
To specify the minimum gap, use BUS<m>:MILStd:MINGap:BITS.
Suffix:
<m>
.
1..4
Parameters:
<MinGapSelect> ON | OFF
*RST: ON
BUS<m>:MILStd:MINGap:BITS <MinGapTime>
Sets a value for the minimum gap.
See also: BUS<m>:MILStd:MINGap:SELect.
Suffix:
<m>
.
1..4
Parameters:
<MinGapTime> Range: 2E-6 to 262E-6
Increment: 100E-9
*RST: 4E-6
Default unit: s
BUS<m>:MILStd:POLarity <Polarity>
Selects the wire on which the bus signal is measured.
Suffix:
<m>
.
1..4
Parameters:
<Polarity> NORMal | INVerted
*RST: NORMal
BUS<m>:MILStd:PRESet <Preset>
Sets the default threshold voltage.
Suffix:
<m>
.
1..4
Parameters:
<Preset> V05 | V2 | V5 | V7 | MANual
*RST: V5
BUS<m>:MILStd:THReshold:HIGH <ThresholdHigh>
Sets the lower threshold level of the signal.
Suffix:
<m>
.
1..4
Protocols