Remote control commands
R&S
®
RTE
1681User Manual 1326.1032.02 ─ 20
Parameters:
<CRCPatternMax> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:HBTO:DATA:DCONdition <DataOperator>
Sets the condition for the data. You can define an exact value or a value range.
Suffix:
<m>
.
1..3
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:HBTO:DATA:
DMIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:HBTO:DATA:DMIN and
TRIGger<m>:HBTO:DATA:DMAX.
*RST:
EQUal
TRIGger<m>:HBTO:DATA:DMIN <DataMin>
Specifies the data bit pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<DataMin> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:HBTO:DATA:DMAX <DataMax>
Sets the end value of the data range if TRIGger<m>:HBTO:DATA:DCONdition is set
to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<DataMax> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
Protocols