Remote control commands
R&S
®
RTE
1708User Manual 1326.1032.02 ─ 20
Parameters:
<StatusBits> Numeric or string pattern, Chapter 18.4.6, "Bit pattern parame-
ter", on page 1076. The string parameter accepts the bit value X
(don't care).
TRIGger<m>:SENT:TDCN <TSDataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<TSDataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one data pattern
to be set with TRIGger<m>:SENT:TDMN.
INRange | OORange
In range / Out of range. To define the range set the minimum
and maximum values with TRIGger<m>:SENT:TDMN and
TRIGger<m>:SENT:TDMX.
*RST: EQUal
TRIGger<m>:SENT:TDMN <TSDataPattern>
Specifies a data pattern, or sets the start value of a data pattern range.
Parameters:
<TSDataPattern> Numeric or string pattern, Chapter 18.4.6, "Bit pattern parame-
ter", on page 1076. The string parameter accepts the bit value X
(don't care).
TRIGger<m>:SENT:TDMX <TSDataPatternTo>
Sets the end value of an identifier range for TRIGger:SENT:TDCN INRange or
OORange.
Parameters:
<TSDataPatternTo> Numeric or string pattern, Chapter 18.4.6, "Bit pattern parame-
ter", on page 1076. The string parameter accepts the bit value X
(don't care).
TRIGger<m>:SENT:STYPe <SMFieldType>
Selects the trigger sequence type for TRIGger:SENT:TYPE SSMSg or ESMSg
(serial message).
Parameters:
<SMFieldType> ID | IDDT
Protocols