Remote control commands
R&S
®
RTE
1722User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:SENT:TDCN <SearchName>,<TSDataOperator>
SEARch:TRIGger:SENT:TDCN? <SearchName>
Sets the operator for a specific data pattern or a data pattern range.
Parameters:
<TSDataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one data pattern
to be set with SEARch:TRIGger:SENT:TDMN.
INRange | OORange
In range / Out of range. To define the range set the minimum
and maximum values with SEARch:TRIGger:SENT:TDMN and
SEARch:TRIGger:SENT:TDMX.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:TDMN <SearchName>,<TSDataPattern>
SEARch:TRIGger:SENT:TDMN? <SearchName>
Specifies a data pattern, or sets the start value of a data pattern range.
Parameters:
<TSDataPattern> Numeric or string pattern, Chapter 18.4.6, "Bit pattern parame-
ter", on page 1076. The string parameter accepts the bit value X
(don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:TDMX <SearchName>,<TSDataPatternTo>
SEARch:TRIGger:SENT:TDMX? <SearchName>
Sets the end value of an identifier range for SEARch:TRIGger:SENT:DCONdition
INRange or OORange.
Parameters:
<TSDataPatternTo> Numeric or string pattern, Chapter 18.4.6, "Bit pattern parame-
ter", on page 1076. The string parameter accepts the bit value X
(don't care).
Parameters for setting and query:
<SearchName>