Remote control commands
R&S
®
RTE
1749User Manual 1326.1032.02 ─ 20
ADVanced
Advanced trigger with TRIGger:CMSB:ADVanced commands.
Only available if frames are defined.
PATTern
Triggers on a data pattern to be specified in TRIGger<m>:
CMSB:PATTern on page 1749
*RST: STARt
TRIGger<m>:CMSB:PATTern <DataPattern>
Sets the pattern match conditions for a payload data check. The trigger is set to the
first occurrence of a matching data bit pattern (which can be freely specified), starting
after the minimum gap time (BUS<m>:CMSB:GAPTime:VALue on page 1742), and
after the detected start of the data frame.
Suffix:
<m>
.
1..3
Parameters:
<DataPattern>
TRIGger<m>:CMSB:ICONdition <DataIdxOperator>
Sets the operator to set a specific bit index (data position).
Suffix:
<m>
.
1..3
Parameters:
<DataIdxOperator> EQUal | GETHan | INRange
EQUal
Equal
GETHan
Greater than or equal
INRange
In range
*RST: INRange
TRIGger<m>:CMSB:IMIN <DataPosition>
Sets the bit index (data position). If TRIGger<m>:CMSB:ICONdition on page 1749
is set to "INRange", "IMIN" sets the start value of the bit index range.
Suffix:
<m>
.
1..3
Parameters:
<DataPosition> Range: 0 to 65535
Increment: 1
*RST: 0
Protocols