Remote control commands
R&S
®
RTE
1760User Manual 1326.1032.02 ─ 20
BUS<m>:CMSB:RESult<n>:CELL<o>:STOP?................................................................ 1763
BUS<m>:CMSB:RESult<n>:CELL<o>:STATe?................................................................1764
BUS<m>:CMSB:RESult<n>:CELL<o>:VALue?................................................................1764
BUS<m>:CMSB:RCOunt?
Returns the count number of decoded result frames in a custom serial bus waveform.
Basically, this is the maximum result index <n> when querying results by using
BUS<m>:CMSB:RESult<n>:XXX.
Suffix:
<m>
.
1..4
Return values:
<Count>
Usage: Query only
BUS<m>:CMSB:RESult<n>:STATe?
Returns the overall state of the frame: either OK or the relevant error condition.
R&S RTE-K50 marks each frame with a status that indicates whether the decode suc-
ceeded or not.
Suffix:
<m>
.
1..4
<n> *
Return values:
<State> OK | LENGth | UNKNown | INComplete | CRC | PARity
OK
The frame was decoded normally and conforms to the frame
description.
LENGth
The length error indicates that the frame ended prematurely, or
an array in the frame had too few elements. The amount of bits
that the software expected (based upon the user's frame
description) was not found before the frame was terminated.
This might occur because a new frame synchronized, or a gap
appeared between the bits.
UNKNown
Unknown error
INComplete
The frame ended prematurely because it extends past the end
of the record.
CRC
Checksum error in cyclic redundancy check (error in data)
PARity
Parity bit error, indicating a transmission error (only available if a
parity is configured)