Remote control commands
R&S
®
RTE
1797User Manual 1326.1032.02 ─ 20
TRIGger<m>:USB:ACONdition <OperatorAddress>
Sets the operator to set a specific address or an address range. The address values
are set with TRIGger<m>:USB:AMIN and TRIGger<m>:USB:AMAX.
Parameters:
<OperatorAddress> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one address
value to be set using TRIGger<m>:USB:AMIN.
INRange | OORange
In range, out of range. These conditions require an address
range to be set using TRIGger<m>:USB:AMIN and
TRIGger<m>:USB:AMAX.
*RST: OFF
TRIGger<m>:USB:AMIN <Address>
Specifies the address, or sets the the start value of an address range. The string
parameter does not accept the bit value X (don't care).
Parameters:
<Address>
TRIGger<m>:USB:AMAX <AddressTo>
Sets the the end value of an address range if TRIGger<m>:USB:ACONdition is set
to INRange or OORange. The string parameter does not accept the bit value X (don't
care).
Parameters:
<AddressTo>
TRIGger<m>:USB:DATA <USBDataType>
Sets the trigger to one out of four different payload data types: DATA0, DATA1, DATA2,
or MDATA.
Parameters:
<USBDataType> D0 | D1 | D2 | MD
D0
Sets the trigger to the DATA0 field (even PID).
D1
Sets the trigger to the DATA1 field (odd PID).
D2
Sets the trigger to the DATA2 field (data packet PID for high-
speed, high bandwidth isochronous transaction in a microframe).