Remote control commands
R&S
®
RTE
1804User Manual 1326.1032.02 ─ 20
TRIGger<m>:USB:TMIN <ET>
Specifies the endpoint type, or sets the the start value of an endpoint type range.
Parameters:
<ET>
TRIGger<m>:USB:TMAX <ETTo>
Sets the the end value of an endpoint type range if TRIGger<m>:USB:TCONdition
is set to INRange or OORange.
Parameters:
<ETTo>
TRIGger<m>:USB:TOKen
<USBTokenType>
Sets the trigger to one out of four different token types: OUT, IN, SOF, or SETup.
Parameters:
<USBTokenType> OUT | IN | SOF | SETup
OUT
Sets the trigger to the OUT token (OUT packet from host to
device).
IN
Sets the trigger to the IN token (IN packet from device to host).
SOF
Sets the trigger to the SOF token (start of frame marker and
frame number).
SETup
Sets the trigger to the SETUP token (address and endpoint
number in OUT transaction for setup to a control pipe).
*RST:
OUT
TRIGger<m>:USB:WADD <AddressCheck>
Defines, whether the address check shall be activated or not.
Parameters:
<AddressCheck> ON | OFF
*RST: OFF
TRIGger<m>:USB:WEND <EndpointCheck>
Defines, whether the endpoint check shall be activated or not.
Parameters:
<EndpointCheck> ON | OFF
*RST: OFF