Measurements
R&S
®
RTE
332User Manual 1326.1032.02 ─ 20
Cursor to center
Sets the vertical cursor line C1 to the center frequency.
Remote command:
CURSor<m>:FFT:TOCenter on page 1247
Center to cursor
Sets the center frequency to the frequency value that is measured at cursor line C1.
Remote command:
CURSor<m>:FFT:SETCenter on page 1247
Threshold
Defines an absolute threshold as an additional condition for the peak search. Only
peaks that exceed the threshold are detected.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
Remote command:
CURSor<m>:THReshold on page 1248
MEASurement<m>:SPECtrum:ATHReshold on page 1272
Peak excursion
Defines a relative threshold, the minimum level value by which the waveform must rise
or fall to be considered as a peak. To avoid identifying noise peaks, enter a peak
excursion value that is higher than the noise levels.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
The following figure shows a cursor measurement on a spectrum waveform:
Cursor measurements