Measurements
R&S
®
RTE
353User Manual 1326.1032.02 ─ 20
Meas. type Symbol Description/Result
Pos. duty cycle R
PosCyc
Positive duty cycle: Width of a positive pulse in relation to the
period in %. The measurement requires at least one complete
period of a triggered signal. Multiple measurements are possi-
ble.
%100
Period
PosPulse
PosCyc
T
T
R
Neg. duty cycle R
NegCyc
Negative duty cycle: Width of a negative pulse in relation to the
period in %. The measurement requires at least one complete
period of a triggered signal. Multiple measurements are possi-
ble.
%100
Period
NegPulse
NegCyc
T
T
R
Delay
Time difference between any two slopes at middle reference
level.
The measurement result is negative if the edge of the second
source comes before the edge of the first source.
See: "Settings for delay and phase measurements (analog sour-
ces)" on page 356
Phase
Phase difference between two waveforms.
Phase = Delay / Period * 360
Burst width
Duration of one burst, measured from the first edge to the last
+
-
Pos. switching T
PosSw
Settling time at rising edges: Time between crossing the lower
reference level and the last return of the signal into the top toler-
ance tube.
+
-
Neg. switching T
NegSw
Settling time at falling edges: Time between crossing the upper
reference level and the last return of the signal into the bottom
tolerance tube. See also "Pos. switching" above.
t?
Pulse train
Duration of N positive pulses, measured from the rising edge of
the first pulse to the falling edge of the N-th pulse. Define N for
the measurement.
Automatic measurements