Data and file management
R&S
®
RTE
486User Manual 1326.1032.02 ─ 20
You can reload exported multiple channels if they are stored in BIN format. The import
asks you to assign each stored waveform to a reference waveform.
Remote command:
EXPort:WAVeform:MULTichannel on page 1402
Selected sources
Select the channels to be included in data export if "Multi-channel export" is enabled.
Waveform1 of up to four input channels can be saved into one file.
Remote command:
CHANnel<m>:EXPortstate on page 1402
Scope
Defines the part of the waveform record that has to be stored.
"Full wave-
form"
Saves the complete waveform record.
"Zoom"
Saves the data included in the zoom area if at least one zoom is
defined for the source waveform. The start and stop values of the
area are shown. If several zooms are defined, select the "Zoom" to be
used for export.
"Cursor"
Saves the data between the cursor lines if at least one cursor mea-
surement is defined for the source waveform. The start and stop val-
ues of the area between the cursor lines are shown. If several cursor
sets are defined, select the "Cursor set" to be used for export.
"Gate"
Saves the data included in the measurement gate if a gated mea-
surement is defined for the source waveform. Select the "Measure-
ment" for which the required gate is defined. The start and stop val-
ues of the gate are shown.
"Manual"
Saves the data between user-defined "Start" and "Stop" values.
Remote command:
EXPort:WAVeform:SCOPe on page 1403
EXPort:WAVeform:STARt on page 1404
EXPort:WAVeform:STOP on page 1404
EXPort:WAVeform:ZOOM on page 1404
EXPort:WAVeform:CURSorset on page 1404
EXPort:WAVeform:MEAS on page 1405
Data logging / Multiple Wfms
The "Data logging / Multiple Wfms" setting enables the export of subsequent acquisi-
tions of the selected waveforms. If "Export history" is disabled, the setting is named
"Data logging", and it exports the data of a running Nx Single acquisition. If "Export his-
tory" is enabled, the setting is named "Multiple Wfms", and it exports the history wave-
form data to file.
Waveform data and results