Protocol analysis
R&S
®
RTE
581User Manual 1326.1032.02 ─ 20
The binary results of data bytes are displayed MSB first.
For CAN protocol, the endianness setting ("Transfer") is a trigger setting and not con-
sidered for decoding.
Figure 13-21: Decoded and binary CAN signal, and decode results
green brackets [...] = Start and end of frame
green frame header = Data frame, ok
cyan frame header = Remote frame, ok
magenta frame header = Overload frame, ok
red frame header = Frame contains an error
no frame header = Error frame
yellow = Identifier
blue = DLC
gray-blue = data
purple = CRC (checksum)
gray = Error frame
red = Error occurred
CAN and CAN FD (option R&S RTE-K3/K9)