Protocol analysis
R&S
®
RTE
594User Manual 1326.1032.02 ─ 20
The "Frame type" is automatically set to "Data", and the "Identifier" is shown for infor-
mation.
Specific settings for the symbolic trigger are:
Message......................................................................................................................594
Trigger on signal..........................................................................................................594
Signal.......................................................................................................................... 594
Condition.....................................................................................................................595
Value, Value min..........................................................................................................595
Value max....................................................................................................................595
Message
Sets the message to be triggered or searched for.
The setting is used by symbolic trigger and symbolic search. Thus, you always trigger
on and search for the same message if symbolic trigger and symbolic search is used at
the same time.
Remote command:
TRIGger<m>:CAN:SYMBolic:MSGValue on page 1525
Trigger on signal
Enables the trigger on a specific signal value that is part of the selected message.
Remote command:
TRIGger<m>:CAN:SYMBolic:TSIGnals on page 1525
Signal
Sets the signal name to be triggered or searched for.
CAN and CAN FD (option R&S
RTE-K3/K9)