EasyManua.ls Logo

Rohde & Schwarz RTE - Page 1194

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1194User Manual 1326.1032.02 ─ 20
TRIGger<m>:TV:POLarity <Polarity>
Sets the polarity of the signal. Note that the sync pulse has the opposite polarity, for
example, a positive signal has a negative sync pulse.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Polarity> POSitive | NEGative
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:TV:LINE
<LineNumber>
Specifies the line number to trigger on. The command is relevant if TRIGger<m>:TV:
MODE is set to LINe.
Usually the lines of the frame are counted beginning from the frame start. For NTSC
signals, the lines are counted per field, not per frame. For these signals, set also the
field with TRIGger<m>:TV:LFIeld.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<LineNumber> Range: Depends on the standard, see table below
Increment: 1
*RST: 1
Usage: Asynchronous command
Standard Minimum value Maximum value
PAL 1 625
PAL-M 1 525
NTSC 1 263 in odd field
262 in even field
SECAM 1 625
480p/60 (P480L60HZ) 1 525
Trigger

Table of Contents

Other manuals for Rohde & Schwarz RTE

Related product manuals