Remote control commands
R&S
®
RTE
1294User Manual 1326.1032.02 ─ 20
MEASurement<m>:VERTical:SCALe <VerticalScale>
Defines the vertical scaling per division, so that the scaling can be adapted automati-
cally during the long term measurement period.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<VerticalScale> Range: 1E-12 to 100E+12
Increment: 10E-6
*RST: 0.5
Default unit: V/div
Usage: Asynchronous command
MEASurement<m>:LTMeas[:STATe]
<ShowDiagram>
Enables long term measurement for a defined number of measurement points (see
MEASurement<m>:LTMeas:COUNt on page 1294) or a specified time
(seeMEASurement<m>:LTMeas:TIME on page 1295).
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<ShowDiagram> ON | OFF
*RST: OFF
Usage: Asynchronous command
MEASurement<m>:LTMeas:COUNt <MeasCnt>
Defines the total number of points to be measured during the long term measurement.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<MeasCnt> Range: 1000 to 200000
Increment: 10
*RST: 1000
Usage: Asynchronous command