Instrument setup
R&S
®
RTE
149User Manual 1326.1032.02 ─ 20
Show labels
If selected, labels mark values on the x- and y-axes in specified intervals in the dia-
gram.
Remote command:
DISPlay:DIAGram:LABels on page 1098
Show div label (x)
If selected, the time scale value is shown at the diagram bottom instead of the horizon-
tal grid labels. For example, 10 ns/div is shown instead of the values 0, 10, 20, 30... ns.
Show fine grid scale
If selected, the crosshair is displayed as a ruler with scale markers. If disabled, the
crosshair is shown as dashed lines.
Remote command:
DISPlay:DIAGram:FINegrid on page 1098
Show crosshair
If selected, a crosshair is displayed in the diagram area. A crosshair allows you to
select a specific data point by its co-ordinates.
Remote command:
DISPlay:DIAGram:CROSshair on page 1098
Keep Y-grid fixed
If enabled, the horizontal grid lines remain in their position when the position of the
curve is changed. Only the values at the grid lines are adapted. Fixed horizontal grid
lines correspond to the behavior of traditional oscilloscopes.
Remote command:
DISPlay:DIAGram:YFIXed on page 1099
Keep X-grid fixed
If enabled, the vertical grid lines remain in their position when the horizontal position is
changed. Only the values at the grid lines are adapted.
Remote command:
DISPlay:DIAGram:XFIXed on page 1099
Show tabs always
If selected, the tab titles of all diagrams are displayed: "Diagram1", "Diagram2" ...
If cleared, the tab titles are not shown except for titles in a tabbed diagram. In tabbed
diagrams, the tab titles are required to change the tabs.
Remote command:
DISPlay:DIAGram:TITLe on page 1098