Remote control commands
R&S
®
RTE
1498User Manual 1326.1032.02 ─ 20
For details, see "Data setup: DLC, NDB, Transfer, Condition, Data min, Data max"
on page 570.
The number of data bytes to be found is set with TRIGger<m>:CAN:DLC.
See also: TRIGger<m>:CAN:BORDer.
Parameters:
<DLCOperator> EQUal | GETHan
For little endian transfer direction, EQUal must be set.
*RST: GETHan
TRIGger<m>:CAN:DLC
<WordCount>
Sets the Data Length Code, the number of data bytes to be found. For complete defini-
tion, set also the operator with TRIGger<m>:CAN:DLCCondition on page 1497.
Parameters:
<WordCount> Range: CAN: 1 to 8, CAN FD: 1 to 15 (64 bytes)
Increment: 1
*RST: 1
TRIGger<m>:CAN:NDBYtes?
Returns the number of data bytes defined by DLC. DLC and NDB are different in CAN
FD for DLCs > 8.
See also: "Data setup: DLC, NDB, Transfer, Condition, Data min, Data max"
on page 570.
Return values:
<NDBytes> Range: 1 to 64
Increment: 1
*RST: 1
Usage: Query only
TRIGger<m>:CAN:FDATa:DPOPerator <DataPosOperator>
Sets the operator to define an exact position or a data range where the instrument
looks for the specified data pattern.
The setting is available in CAN FD option R&S RTE-K9.
The position can be defined if the data field of the frame is longer than 8 bytes - if
TRIGger<m>:CAN:DLC≥9.
Parameters:
<DataPosOperator> ANY | OFF | EQUal | GETHan | INRange | RANGe
ANY = OFF
The data position is not relevant for the trigger condition.
Protocols