Remote control commands
R&S
®
RTE
1499User Manual 1326.1032.02 ─ 20
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with TRIGger<m>:CAN:FDATa:DPOSition.
INRange = RANGe
In range: Set the minimum and maximum value of the range with
TRIGger<m>:CAN:FDATa:DPOSition and TRIGger<m>:
CAN:FDATa:DPTO.
*RST: ANY
TRIGger<m>:CAN:FDATa:DPOSition <DataPosition>
Defines the number of the first data byte at which the data pattern may start.
The setting is available in CAN FD option R&S RTE-K9.
Parameters:
<DataPosition> Range: 1 to 57
Increment: 1
*RST: 1
TRIGger<m>:CAN:FDATa:DPTO <DataPositionTo>
Sets the number of the last byte at which the required data pattern may start.
Parameters:
<DataPositionTo> Range: 8 to 64
Increment: 1
*RST: 8
TRIGger<m>:CAN:ACKerror <AckError>
Triggers when the transmitter does not receive an acknowledgment - a dominant bit
during the Ack Slot.
The trigger type has to be set before: TRIGger<m>:CAN:TYPE to ERRC.
Parameters:
<AckError> ON | OFF
*RST: ON
TRIGger<m>:CAN:BITSterror <BitStuffError>
Triggers if astuff error occurs - when the 6th consecutive equal bit level in the men-
tioned fields is detected.
The trigger type has to be set before: TRIGger<m>:CAN:TYPE to ERRC.
Parameters:
<BitStuffError> ON | OFF
*RST: ON