Remote control commands
R&S
®
RTE
1512User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:CAN:ITYPe <SearchName>,<IdentifierType>
SEARch:TRIGger:CAN:ITYPe? <SearchName>
Selects the format of data and remote frames: 11 bit for CAN base frames, or 29 bits
for CAN extended frames.
Parameters:
<IdentifierType> ANY | B11 | B29
*RST: ANY
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CAN:ICONdition
<SearchName>,<IdOperator>
SEARch:TRIGger:CAN:ICONdition? <SearchName>
Sets the operator to set a specific identifier or an identifier range.
Parameters:
<IdOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These condition require one identifier pat-
tern to be set with SEARch:TRIGger:CAN:IMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with SEARch:TRIGger:CAN:IMIN and SEARch:
TRIGger:CAN:IMAX.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CAN:IMAX <SearchName>,<IdPatternTo>
SEARch:TRIGger:CAN:IMAX? <SearchName>
Sets the the end value of an identifier range if SEARch:TRIGger:CAN:ICONdition
is set to INRange or OORange.
Parameters:
<IdPatternTo> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.
Parameters for setting and query:
<SearchName>
Protocols