Remote control commands
R&S
®
RTE
1514User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:CAN:DMAX <SearchName>,<DataPatternTo>
SEARch:TRIGger:CAN:DMAX? <SearchName>
Sets the the end value of an data range if SEARch:TRIGger:CAN:DCONdition is
set to INRange or OORange.
Parameters:
<DataPatternTo> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CAN:DLCCondition <SearchName>,<DLCOperator>
SEARch:TRIGger:CAN:DLCCondition? <SearchName>
Operator to set the data length code for search.
Parameters:
<DLCOperator> EQUal | GETHan
*RST: GETHan
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CAN:DLC <SearchName>,<WordCount>
SEARch:TRIGger:CAN:DLC? <SearchName>
Sets the Data Length Code, the number of data bytes to be found. For complete defini-
tion, set also the operator with SEARch:TRIGger:CAN:DLCCondition.
Parameters:
<WordCount> Range: 1 to 8
Increment: 1
*RST: 1
Parameters for setting and query:
<SearchName>
SEARch:RESult:CAN:FRAMe<m>:NDBYtes? <SearchName>
Returns the number of data bytes defined by DLC. DLC and NDB are different in CAN
FD for DLCs > 8.
See also: "Data setup: DLC, NDB, Transfer, Condition, Data min, Data max"
on page 570.