Parameters:
<DataPositionTo> Range:  0  to  255
Increment:  1
*RST:  0
TRIGger<m>:FLXRay:DCONdition <DataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
OFF = ANY
The data position is not relevant for the trigger condition.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data posi-
tion to be set with TRIGger<m>:FLXRay:DMIN.
INRange | OORange
In range, Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:FLXRay:DMIN and TRIGger<m>:
FLXRay:DMAX.
*RST: 
EQUal
TRIGger<m>:FLXRay:DMIN <DataPattern>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:FLXRay:DMAX <DataPatternTo>
Sets the the end value of an data range if the operator TRIGger<m>:FLXRay:
DCONdition is set to INRange or OORange.
Parameters:
<DataPatternTo> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:FLXRay:SYMBol <Symbol>
Triggers on a symbol or on a wakeup pattern.
Parameters:
<Symbol> CASMts | WAKeup