Remote control commands
R&S
®
RTE
1678User Manual 1326.1032.02 ─ 20
TRIGger<m>:HBTO:DADDress:MAX <DestAddrPattMax>
Sets the end value of the destination address range if TRIGger<m>:HBTO:
DADDress:CONDition is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<DestAddrPattMax> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:HBTO:SADDress:CONDition
<SrcAddrOperator>
Sets the condition for the source address. You can define an exact value or a value
range.
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:HBTO:
SADDress:MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:HBTO:SADDress:MIN and
TRIGger<m>:HBTO:SADDress:MAX.
*RST:
EQUal
TRIGger<m>:HBTO:SADDress:MIN <SrcAddrPattMin>
Specifies the source address bit pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrPattMin> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
Protocols