Remote control commands
R&S
®
RTE
1800User Manual 1326.1032.02 ─ 20
SE1error
Triggers on the illegal bus state Single Ended 1 (SE1 = both
lines high).
GLITcherr
Triggers on any glitching error (illegal bit period, see USB stan-
dard for the definition of glitching).
*RST: ANY
TRIGger<m>:USB:FCONdition <FrameNoOperator>
Sets the operator to set a specific frame number or a frame number range. The frame
number values are set with TRIGger<m>:USB:FMIN and TRIGger<m>:USB:FMAX.
Parameters:
<FrameNoOperator> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one frame num-
ber value to be set using TRIGger<m>:USB:FMIN.
INRange | OORange
In range, out of range. These conditions require a range of
frame number values to be set using TRIGger<m>:USB:FMIN
and TRIGger<m>:USB:FMAX.
*RST: OFF
TRIGger<m>:USB:FMIN <FrameNumber>
Specifies the frame number, or sets the the start value of a frame number range.
Parameters:
<FrameNumber>
TRIGger<m>:USB:FMAX <FrameNumberTo>
Sets the the end value of a frame number range if TRIGger<m>:USB:FCONdition is
set to INRange or OORange.
Parameters:
<FrameNumberTo>
TRIGger<m>:USB:HAND <HandshakeType>
Sets the trigger to one out of four different handshake types: ACK, NAK, STALl, or
NYET.
Parameters:
<HandshakeType> ACK | NAK | STALl | NYET