Remote control commands
R&S
®
RTE
1896User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:CXPI:TYPE <SearchName>,<Type>
SEARch:TRIGger:CXPI:TYPE? <SearchName>
Searches for the selected CXPI frame type.
Parameters:
<Type> STARt | NORMal | NPOLl | SLEEP | LONG | LPOLl | PID |
PTYPe | PTPD | ERRor
STARt
Triggers on the STARt of the frame.
NORMal
Triggers on a NORMal frame. Optionally you can define a frame
ID condition, a NM, a CT, a DLC condition, a data pattern condi-
tion and an index condition.
NPOLl
Triggers on a Normal POLL frame. Optionally you can define a
frame ID condition, a NM, a CT, a DLC condition, a data pattern
condition and an index condition.
SLEEP
Triggers on a SLEEP frame. Optionally you can define a NM, a
CT, a data pattern condition and an index condition.
LONG
Triggers on a Long frame. Optionally you can define a frame ID
condition, a NM, a CT, a DLC extend condition, a data pattern
condition and an index condition.
LPOLl
Triggers on a Long POLL frame.
PID
Triggers on a Protected ID field. Optionally, you can define the
frame ID pattern.
PTYPe
Triggers on a Protected TYPE field.
PTPID
Triggers on a PType field followed by a PID field.Optionally, you
can define the frame ID pattern.
ERRor
Trigger on selected error types. Define the error types with:
SEARch:TRIGger:CXPI:ERRor:CRC on page 1893
SEARch:TRIGger:CXPI:ERRor:CRC on page 1893
SEARch:TRIGger:CXPI:ERRor:DLC on page 1893
SEARch:TRIGger:CXPI:ERRor:IBS on page 1893
SEARch:TRIGger:CXPI:ERRor:IFS on page 1893
SEARch:TRIGger:CXPI:ERRor:LENGth on page 1894
SEARch:TRIGger:CXPI:ERRor:PARity on page 1894
SEARch:TRIGger:CXPI:ERRor:UART on page 1894
*RST:
STARt
Protocols