Remote control commands
R&S
®
RTE
1930User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..2
Parameters:
<Frequency1> Range: 1E-3 to 100E+6
Increment: 1
*RST: 1E+6
Default unit: Hz
WGENerator<m>:MODulation:FSK:FTWO <Frequency2>
Sets the frequency of the modulated waveform, if WGENerator<m>:MODulation:
TYPE is set to FSK.
Suffix:
<m>
.
1..2
Parameters:
<Frequency2> Range: 1E-3 to 100E+6
Increment: 1
*RST: 1000
Default unit: Hz
WGENerator<m>:MODulation:FSK[:RATE] <Rate>
Sets the hop rate, the time before a switch from the carrier frequency set with
WGENerator<m>:MODulation:FSK:FONE and the modulation frequency set with
WGENerator<m>:MODulation:FSK:FTWO.
Suffix:
<m>
.
1..2
Parameters:
<Rate> Range: 1E-3 to 1E+6
Increment: 1
*RST: 1000
Default unit: Hz
WGENerator<m>:MODulation:PWM:DCYCle <SquareDutyCycle>
Sets the duty cycle of a square waveform for pulse width modulation, if
WGENerator<m>:MODulation:TYPE is set to PWM.
Suffix:
<m>
.
1..2
Parameters:
<SquareDutyCycle> Range: 10 to 90
Increment: 1
*RST: 50
Default unit: %
Waveform generator (option R&S RTE-B6)