Triggers
R&S
®
RTE
276User Manual 1326.1032.02 ─ 20
Ground
If the selected trigger source is the external trigger input, you can connect the trigger
input to the ground.
Remote command:
TRIGger<m>:ANEDge:GND on page 1170
Coupling
You can set the coupling in the trigger configuration.
"DC 50 Ω"
Direct connection with 50 Ω termination, passes both DC and AC
components of the trigger signal.
"DC 1 MΩ"
Direct connection with 1 MΩ termination, passes both DC and AC
components of the trigger signal.
"AC"
Connection through capacitor, removes unwanted DC and very low-
frequency components.
Remote command:
TRIGger<m>:ANEDge:COUPling on page 1168
Filter
If the selected trigger source is "Extern" (external trigger input), you can directly select
a filter to reject high or low frequencies.
For all other trigger sources, you can add a digital filter using the Digital Filter Setup.
"Off"
The trigger signal is not filtered.
"Highpass"
Frequencies below the "Cut-off" frequency are rejected, higher fre-
quencies pass the filter.
You can adjust the "Cut-off" frequency, the default is 50 kHz.
"Lowpass"
Frequencies higher than the "Cut-off" frequency are rejected, lower
frequencies pass the filter.
You can adjust the "Cut-off" frequency, the default is 50 kHz.
Remote command:
TRIGger<m>:ANEDge:FILTer on page 1169
TRIGger<m>:ANEDge:CUToff:HIGHpass on page 1169
TRIGger<m>:ANEDge:CUToff:LOWPass on page 1170
External trigger input