Remote control commands
R&S
®
RTE
1614User Manual 1326.1032.02 ─ 20
TRIGger<m>:MILStd:CDST:ICONdition <InfoOperator>
Sets the operator to set a specific info for the 9th to 19th bit of a command or status
word.
Parameters:
<InfoOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set with TRIGger<m>:MILStd:CDST:IMIN.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:MILStd:CDST:IMIN and
TRIGger<m>:MILStd:CDST:IMAX.
*RST: EQUal
TRIGger<m>:MILStd:CDST:IMAX <InfoPatternMax>
Sets the end value of the info range if TRIGger<m>:MILStd:CDST:ICONdition is
set to INRange.
Parameters:
<InfoPatternMax> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:CDST:IMIN <InfoPatternMin>
Specifies an info or sets the the start value of an info range.
Parameters:
<InfoPatternMin> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:DATA:DCONdition <DataOperator>
Sets the operator to set a specific data pattern.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set with TRIGger<m>:MILStd:DATA:DMIN.
Protocols