Remote control commands
R&S
®
RTE
1617User Manual 1326.1032.02 ─ 20
Parameters:
<MaxResponseTime> Range: 2E-6 to 262E-6
Increment: 100E-9
*RST: 14E-6
Default unit: s
TRIGger<m>:MILStd:MAXResponse:SELect <MaxRespSel>
Enables the trigger on exceeding the maximum response time if TRIGger<m>:
MILStd:TYPE is set to ERRor. You can set the maximum time with: TRIGger<m>:
MILStd:MAXResponse:BITS.
Parameters:
<MaxRespSel> ON | OFF
*RST: ON
TRIGger<m>:MILStd:MINGap:BITS
<MinGapTime>
Sets the value for the minimum gap to be triggered on if TRIGger<m>:MILStd:TYPE
is set to ERRor.
Parameters:
<MinGapTime> Range: 2E-6 to 262E-6
Increment: 100E-9
*RST: 4E-6
Default unit: s
TRIGger<m>:MILStd:MINGap:SELect <MinGapSelect>
Enables triggering when the minimum gap is out of range if TRIGger<m>:MILStd:
TYPE is set to ERRor. You can set the minimum gap with: TRIGger<m>:MILStd:
MINGap:BITS.
Parameters:
<MinGapSelect> ON | OFF
*RST: ON
TRIGger<m>:MILStd:STATus:BCReceived <BroadcastCmd>
Triggers on the state of the broadcast command received bit of the status word if
TRIGger<m>:MILStd:TYPE is set to STATword.
Parameters:
<BroadcastCmd> ONE | ZERO | DC
*RST: DC
Protocols