Acquisition and waveform setup
R&S
®
RTE
198User Manual 1326.1032.02 ─ 20
"Run Continuous"
Is the default assignment. Starts or stops the acquisition (same as
RUN CONT key).
"Run single"
Starts a defined number of acquisitions (same as [RUN N× SINGLE]
key).
"Auto set"
Starts the autoset procedure (same as AUTOSET key).
"AutoZero"
Starts an auto zero measurement, see "Detect AutoZero, Use Auto-
Zero" on page 195.
"Set offset to mean"
Performs an automatic compensation for a DC component of the
input signal using the result of a background mean measurement.
See: "Offset to mean" on page 211.
"Print"
Prints the current display according to the "Printer control" settings in
the "Print" dialog box. Depending on the selected printer, you can
print to a local or network driver, or save to a file. See also: Chap-
ter 12.4, "Screenshots", on page 501.
"Save image to file"
Saves the current display as image according to the image settings.
See also Chapter 12.4, "Screenshots", on page 501.
"No action"
Select this option to prevent unwanted actions due to unintended
usage of the micro button.
"Find trigger
level"
Sets the trigger level automatically to 0.5 * (MaxPeak – MinPeak).
"Create report"
Creates and saves a report using the report settings. See also Chap-
ter 12.5.1, "Report settings", on page 508.
"Probe mode"
Only available if a R&S RT-ZM modular probe is connected. Sets the
measurement mode of the modular probe. See also "Probe mode"
on page 207.
"Probe Setup"
Opens the "Probes Setup" dialog box.
Remote command:
PROBe<m>:SETup:MODE on page 1127
ProbeMeter
The integrated R&S ProbeMeter of active voltage probes with Rohde & Schwarz probe
interface is a voltmeter. It measures DC voltages between the probe tip and ground
connection or between the probe tips with very high precision. The R&S ProbeMeter
enables ground-referenced measurements of voltages. The measurement is performed
continuously and in parallel to the measurements of the oscilloscope.
●
"Probemeter"
Select "Probemeter" to activate the integrated R&S ProbeMeter of active R&S
probes. The measured voltages are displayed in the "ProbeMeter" result box on
the screen.
●
ProbeMeter measurement results of single-ended active R&S probes
Measures the voltage between the probe tip and the ground.
Probes