Acquisition and waveform setup
R&S
®
RTE
199User Manual 1326.1032.02 ─ 20
●
ProbeMeter measurement results of differential and modular R&S probes
You can select the voltage to be measured by the differential active probe:
– "Differential / Common Mode":
Differential voltage is the voltage between the positive and negative signal
sockets.
Common mode voltage is the mean voltage between the signal sockets and
the ground socket. It measures the voltage level relative to ground, for exam-
ple, to check the operating voltage window.
– "Single Ended Pos/Neg": Measures the voltage between the positive/negative
signal socket and the ground.
The ProbeMeter always measures the common mode and differential voltages.
Single-ended voltages are calculated values:
V
p
= V
cm
+ 0.5 * V
in
and V
n
= V
cm
- 0.5 * V
in
Remote command:
Channel probes:
PROBe<m>:PMETer:VISibility on page 1128
PROBe<m>:SETup:DISPlaydiff on page 1128
PROBe<m>:PMETer:RESults:SINGle? on page 1129
PROBe<m>:PMETer:RESults:POSitive? on page 1130
PROBe<m>:PMETer:RESults:NEGative? on page 1130
PROBe<m>:PMETer:RESults:DIFFerential? on page 1129
PROBe<m>:PMETer:RESults:COMMon? on page 1129
Probe on trigger input:
TRPRobe:SETup:DISPlaydiff on page 1128
TRPRobe:PMETer:VISibility on page 1128
TRPRobe:PMETer:RESults:SINGle? on page 1129
TRPRobe:PMETer:RESults:POSitive? on page 1130
Probes