Measurements
R&S
®
RTE
343User Manual 1326.1032.02 ─ 20
(Relative) Start
Defines the starting value for the gate.
Remote command:
CALCulate:MATH<m>:FFT:GATE:ABSolute:STARt on page 1322
CALCulate:MATH<m>:FFT:GATE:RELative:STARt on page 1323
MEASurement<m>:GATE:ABSolute:STARt on page 1303
MEASurement<m>:GATE:RELative:STARt on page 1303
SEARch:GATE:ABSolute:STARt on page 1380
SEARch:GATE:RELative:STARt on page 1380
(Relative) Stop
Defines the end value for the gate.
Remote command:
CALCulate:MATH<m>:FFT:GATE:ABSolute:STOP on page 1323
CALCulate:MATH<m>:FFT:GATE:RELative:STOP on page 1323
MEASurement<m>:GATE:ABSolute:STOP on page 1303
MEASurement<m>:GATE:RELative:STOP on page 1303
SEARch:GATE:ABSolute:STOP on page 1380
SEARch:GATE:RELative:STOP on page 1381
Zoom coupling
Zoom coupling is available if a zoom is defined. As long as "Zoom coupling" is
enabled, the gate area is defined identically to the zoom area - if you change the zoom,
the gate changes as well.
If several zoom diagrams are defined, select the zoom diagram to be used for gating.
The "Start" and "Stop" values of the gate are adjusted accordingly.
Zoom coupling can be set for measurement gates, FFT gates, and search gates. The
zoom must be defined on the diagram that contains the signal source of the measure-
ment, FFT, or search.
Remote command:
MEASurement<m>:GATE:ZCOupling on page 1304
MEASurement<m>:GATE:ZDIagram on page 1305
CALCulate:MATH<m>:FFT:GATE:ZCOupling on page 1324
SEARch:GATE:ZCOupling on page 1381
SEARch:GATE:ZDIagram on page 1381
Cursor coupling
If enabled, the gate area is defined by the cursor lines of an active cursor measure-
ment.
If several cursor measurements are enabled, select the cursor set to be used for gat-
ing. The "Start" and "Stop" values of the gate are adjusted to the values of the cursor
line positions. The measurement is limited to the part of the waveform between the cur-
sor lines.
Remote command:
MEASurement<m>:GATE:CCOupling on page 1304
MEASurement<m>:GATE:CURSor on page 1304
Automatic measurements