1004 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
32 :TIMebase Commands
:TIMebase:WINDow:RANGe
(see page 1276)
Command Syntax
:TIMebase:WINDow:RANGe <range value>
<range value> ::= range value in seconds in NR3 format
The :TIMebase:WINDow:RANGe command sets the full-scale horizontal time in
seconds for the zoomed (delayed) window. The range is 10 times the current
zoomed view window seconds per division setting. The main sweep range
determines the range for this command. The maximum value is one half of the
:TIMebase:RANGe value.
Query Syntax
:TIMebase:WINDow:RANGe?
The :TIMebase:WINDow:RANGe? query returns the current window timebase
range setting.
Return Format
<value><NL>
<value> ::= range value in seconds
See Also • "Introduction to :TIMebase Commands" on page 996
• ":TIMebase:RANGe" on page 999
• ":TIMebase:POSition" on page 998
• ":TIMebase:SCALe" on page 1001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals