:MEASure Power Commands 23
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 533
:MEASure:EFFiciency
(see page 1276)
Command Syntax
:MEASure:EFFiciency
The :MEASure:EFFiciency command installs an efficiency (output power / input
power) measurement on screen.
Before sending this command or query, you must specify the channels probing the
input voltage, input current, output voltage, and output current (using the
:POWer:SIGNals:SOURce:VOLTage<i> and :POWer:SIGNals:SOURce:CURRent<i>
commands) and you must perform the automated signals setup (using the
:POWer:SIGNals:AUTosetup EFFiciency command).
Query Syntax
:MEASure:EFFiciency?
The :MEASure:EFFiciency query returns the measured efficiency as a percent
value.
Return Format
<return_value><NL>
<return_value> ::= percent value in NR3 format
See Also • ":POWer:SIGNals:SOURce:VOLTage<i>" on page 642
• ":POWer:SIGNals:SOURce:CURRent<i>" on page 641
• ":POWer:SIGNals:AUTosetup" on page 624
• ":POWer:EFFiciency:APPLy" on page 595

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals