:MEASure Commands 22
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 501
:MEASure:STATistics
(see page 1276)
Command Syntax
:MEASure:STATistics <type>
<type> ::= {{ON | 1} | CURRent | MINimum | MAXimum | MEAN | STDDev
| COUNt}
The :MEASure:STATistics command determines the type of information returned by
the :MEASure:RESults? query. ON means all the statistics are on.
Query Syntax
:MEASure:STATistics?
The :MEASure:STATistics? query returns the current statistics mode.
Return Format
<type><NL>
<type> ::= {ON | CURR | MIN | MAX | MEAN | STDD | COUN}
See Also • "Introduction to :MEASure Commands" on page 458
• ":MEASure:RESults" on page 493
• ":MEASure:STATistics:DISPlay" on page 502
• ":MEASure:STATistics:RESet" on page 505
• ":MEASure:STATistics:INCRement" on page 503
Example Code • "Example Code" on page 493