26 :POWer Commands
:POWer:HARMonics:LINE
(see page 1276)
Command Syntax
:POWer:HARMonics:LINE <frequency>
<frequency> ::= {F50 | F60 | F400}
The :POWer:HARMonics:LINE command specifies the line frequency setting for the
current harmonics analysis:
• F50 — 50 Hz.
• F60 — 60 Hz.
• F400 — 400 Hz.
Query Syntax
:POWer:HARMonics:LINE?
The :POWer:HARMonics:LINE query returns the line frequency setting.
Return Format
<frequency><NL>
<frequency> ::= {F50 | F60 | F400}