EasyManuals Logo

Keysight Technologies InfiniiVision 3000T X Series User Manual

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #602 background imageLoading...
Page #602 background image
602 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
26 :POWer Commands
:POWer:HARMonics:LINE
(see page 1276)
Command Syntax
:POWer:HARMonics:LINE <frequency>
<frequency> ::= {F50 | F60 | F400}
The :POWer:HARMonics:LINE command specifies the line frequency setting for the
current harmonics analysis:
• F50 — 50 Hz.
• F60 — 60 Hz.
• F400 — 400 Hz.
Query Syntax
:POWer:HARMonics:LINE?
The :POWer:HARMonics:LINE query returns the line frequency setting.
Return Format
<frequency><NL>
<frequency> ::= {F50 | F60 | F400}
See Also • ":POWer:HARMonics:APPLy" on page 598
• ":POWer:HARMonics:DATA" on page 599
• ":POWer:HARMonics:DISPlay" on page 600
• ":POWer:HARMonics:FAILcount" on page 601
• ":POWer:HARMonics:POWerfactor" on page 603
• ":POWer:HARMonics:RUNCount" on page 604
• ":POWer:HARMonics:STANdard" on page 605
• ":POWer:HARMonics:STATus" on page 606
• ":POWer:HARMonics:THD" on page 607

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals