Keysight InfiniiVision 3000T X-Series Oscilloscopes
Programmer's Guide
9 :CALibrate Commands
Utility commands for viewing calibration status and for starting the user
calibration procedure. See "Introduction to :CALibrate Commands" on page 256.
Table 87 :CALibrate Commands Summary
Command Query Options and Query Returns
n/a :CALibrate:DATE? (see
page 257)
<return value> ::=
<year>,<month>,<day>; all in NR1
format
:CALibrate:LABel
<string> (see
page 258)
:CALibrate:LABel?
(see page 258)
<string> ::= quoted ASCII string
up to 32 characters
:CALibrate:OUTPut
<signal> (see
page 259)
:CALibrate:OUTPut?
(see page 259)
<signal> ::= {TRIGgers | MASK |
WAVEgen | WGEN1 | WGEN2}
Note: WAVE and WGEN1 are
equivalent.
Note: WGEN2 only available on
models with 2 WaveGen outputs.
n/a :CALibrate:PROTected?
(see page 260)
{PROTected | UNPRotected}
:CALibrate:STARt (see
page 261)
n/a n/a
n/a :CALibrate:STATus?
(see page 262)
<return value> ::=
<status_code>,<status_string>
<status_code> ::= an integer
status code
<status_string> ::= an ASCII
status string
n/a :CALibrate:TEMPeratur
e? (see page 263)
<return value> ::= degrees C
delta since last cal in NR3
format
n/a :CALibrate:TIME? (see
page 264)
<return value> ::=
<hours>,<minutes>,<seconds>; all
in NR1 format