:TRIGger Commands 33
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1053
:TRIGger:PATTern:LESSthan
(see page 1276)
Command Syntax
:TRIGger:PATTern:LESSthan <less_than_time>[<suffix>]
<less_than_time> ::= maximum trigger duration in seconds
in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:PATTern:LESSthan command sets the maximum duration for the
defined pattern when :TRIGger:PATTern:QUALifier is set to LESSthan.
Query Syntax
:TRIGger:PATTern:LESSthan?
The :TRIGger:PATTern:LESSthan? query returns the duration time for the defined
pattern.
Return Format
<less_than_time><NL>
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":TRIGger:PATTern" on page 1049
• ":TRIGger:PATTern:QUALifier" on page 1054
• ":TRIGger:MODE" on page 1016