:SBUS<n> Commands 29
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 733
:SBUS<n>:CAN:SIGNal:FDBaudrate
(see page 1276)
Command Syntax
:SBUS<n>:CAN:SIGNal:FDBaudrate <baudrate>
<baudrate> ::= integer from 10000 to 10000000 in 100 b/s increments.
The :SBUS<n>:CAN:SIGNal:FDBaudrate command sets the CAN FD baud rate
from 10 kb/s to 10 Mb/s in 100 b/s increments. If you enter a baud rate that is not
divisible by 100 b/s, the baud rate is set to the nearest baud rate divisible by
100 b/s.
For CAN FD, both the standard rate settings (see
:SBUS<n>:CAN:SIGNal:BAUDrate) and the FD rate settings must be set correctly;
otherwise, false triggers may occur.
Query Syntax
:SBUS<n>:CAN:SIGNal:FDBaudrate?
The :SBUS<n>:CAN:SIGNal:FDBaudrate? query returns the current CAN FD baud
rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 10000 to 10000000 in 100 b/s increments.
See Also ":SBUS<n>:CAN:FDSPoint" on page 729
":SBUS<n>:CAN:SIGNal:BAUDrate" on page 731