504 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
22 :MEASure Commands
:MEASure:STATistics:MCOunt
(see page 1276)
Command Syntax
:MEASure:STATistics:MCOunt <setting>
<setting> ::= {INFinite | <count>}
<count> ::= 2 to 2000 in NR1 format
The :MEASure:STATistics:MCOunt command specifies the maximum number of
values used when calculating measurement statistics.
Query Syntax
:MEASure:STATistics:MCOunt?
The :MEASure:STATistics:MCOunt? query returns the current measurement
statistics max count setting.
Return Format
<setting><NL>
<setting> ::= {INF | <count>}
<count> ::= 2 to 2000
See Also "Introduction to :MEASure Commands" on page 458
":MEASure:RESults" on page 493
":MEASure:STATistics" on page 501
":MEASure:STATistics:DISPlay" on page 502
":MEASure:STATistics:RSDeviation" on page 506
":MEASure:STATistics:RESet" on page 505
":MEASure:STATistics:INCRement" on page 503